Model 9130 Triple Output Programmable DC Power Supply INSTRUCTION MANUAL MODEL 9130
8displayed current value (actual output value) is the same as the set value. 7) Ensure that the current can be adjusted from zero to the maximum rat
97. To Rack Mount the Instrument You can mount the power supply in a standard 19-inch rack cabinet using the IT-E151 rack mount kit. Note: Remove th
103. Specifications 3.1 Specifications Parameter 9130 Voltage 0~30V×2,0~5V×1 Current 0~3A×2,0~3A×1 Output ratings LVP* 0~31V×2,0~6V×1 Voltage ≤0
113.2 Supplemental Characteristics State Storage Memory 50 user-configurable stored states Recommended Calibration Interval 1 year AC Input Ratings
124. Front-panel Operation So far we have covered the quick start chapter which briefly introduced the front panel operation and how to check basic v
134.2 Panel Description 41 27536Esc0●98RecallMenuEnterV-setV-setI-setI-setSaveOn/Off 0~9 Use keys 1~3 to control the o
14scroll through the complete menu listed below. Press Enter to enter the selected menu function, press Esc to return to the previous menu. A ”↑↓” s
154.5 Panel Operation Channel Operation When the power supply is in “METER” mode, you can select a channel using the arrow keys . OUT ON/OFF Pressi
16adjust the first channel, press 5 to control the second channel and press 6 to control the third channel). For example, to set voltage for the fi
17Over Temp 4.6 Menu Function In menu operation, the keys or the knob can be used to select the menu. Enter is used to confirm a selection. The E
Safety information Please review the following safety precautions before operating our equipment. General information The following safety precautio
18back to meter mode or output mode The minimum is 4 seconds, and the maximum is 9999 seconds. Enter a numerical value followed by Enteror press
19Connection diagram example: Channels 1 and 2 connected in series Out Parallel Set This function configures the instrument for parallel operat
20Connection diagram example: Channels 1 and 2 connected in parallel Note: When configuring for series or parallel operation, the user must wi
21Power Model: the model of power supply For example: 30V,3A*2CH 5V,3A*1CH Power SN: the serial number For example: 001001156074001165 Soft Version:
225. Remote Operation The DB9 interface connector on the rear panel of the power supply can be connected to a RS-232 or USB interface via the include
235.2 Communication between Power Supply and PC Before putting the instrument into remote operation, make sure that the baud rate, parity bit and com
245.3 SCPI Command Overview 5.3.1 Common IEEE488.2 commands *CLS *ESE *ESE? *ESR? *IDN? *OPC *OPC? *PSC *PSC? *RST *SRE *SRE? *STB? *S
25 INSTrument [:SELect] {FIRst|SECOnd|THIrd} [:SELect]? NSELect {1|2|3} NSELect? OUTPut [:STATe] {0|1} [:STATe]? [SOURce:] CU
26 DISPlay: [:WINDow][:STATe] {OFF|ON} [:WINDow][:STATe]? MEASure[:SCALer] :CURRent[:DC]? [:VOLTage][:DC]? :POWer[:DC]? 5.4 SCPI Comma
27 The Structure of the condition register is as followed: condition event enable condition event enable
1TABLE OF CONTENTS 1. INTRODUCTION...
285.4.2 SCPI Command Description Common IEEE488.2 Commands *CLS This command clears the following registers: Standard event status register
29 For example:BK PRECISION, 9130, 000000000000111101,V1.68 *OPC When all commands sent to the instrument prior to this command have been execut
30Command syntax:*SAV Parameter:0~49 Example:*SAV 10 *RCL This command recalls parameter saved previously (using the *SAV command). Com
31 SYSTem:ADDRess This command reads the communication address of the power supply. Command syntax:SYSTem:ADDRess? Example:SYST:ADDR? ST
32 Reset value:consult PSC command Example:STAT:OPER:ENAB 110 STATus:OPERation:ENABle? This command queries parameters from the operation enab
33the operation condition sub register is reset. Note: this command is only valid for current channel. Command syntax:STATus:OPERation:INSTrument
34 Example:OUTPut:TIMer:DATA 100 OUTPut:TIMer:DATA? This command queries the time of the output-timer for the current channel. Command synta
35 [SOURce:]VOLTage:PROTection[:LEVel][:IMMediate][:AMPLitude] This command sets the max. voltage for the current channel. Command syntax:[SOURce
36 CALibration:SECure[:STATe] This command sets the protection mode to “enable” or “disable”. The power supply will be calibrated only if the ca
37This command sets the new calibration password. Command syntax:CALibration:SECure:CODE <parameter> Parameter:the password (1 through 4digit
26. PV9130 SOFTWARE GUIDE... 38 6.1 Introduction...
386. PV9130 Software Guide 6.1 Introduction PV9130 is a software application for the 9130 Triple Output Programmable DC Power Supply. To use this
391 Configure the software operating parameters and serial communication parameters. 2 Voltage chart 3 Voltage and current displayed for all 3 c
406.3.2 Chart Description The voltage and current chart can help you to analyze voltage and current changes more easily. If you want to observe the
416.4 Operation 6.4.1 Configure the system The first step is to configure the system and set up communication. Click the button and the followi
426.5 Test sequence generation 6.5.1 Quickset Hot Keys Hotkey:Right-click each Hotkey Button, it will display , Click on it, it will display as foll
43 6.5.2 Go/NG This tool allows you to write a simple test sequence including min/max values. The test program generates a pass fail verdict and a
44 [left intentionally blank]
Service Information Warranty Service: Please return the product in the original packaging with proof of purchase to the address below. Clearly stat
22820 Savi Ranch Parkway Yorba Linda, CA 92887 www.bkprecision.com © 2007 B&K Precision Corp.
31. Introduction Description The 9130 is a fully programmable triple Output DC Power Supply delivering 0-30V/0-3A on 2 outputs and 0-5V/0-3A on 1 ou
42. Quick Start This chapter will help you getting acquainted with the front panel operation of this power supply. 2.1 Front & Rear Panel Front p
5Rear Panel 1 2 3 4 5 6 1) Terminal block for GND referenc
62.Connect the power cord and turn on the power supply When you turn on the power supply, the front-panel display will light up briefly while the pow
7CH X Lost Calibration... Note: “X” denotes the channel for which calibration data was lost. If the calibration data in the EEPROM is