SCPI Command Reference
72
:CALCulate[1] Subsystem
Use this subsystem for commands to configure and control MXB and percent math calculations.
:FORMat <name>
Command syntax:
:CALCulate[1]:FORMat <name>
Command Parameter:
<name> = NONE No calculations
MXB MX+B math calculation
PERCent Percent math calculation
Query:
:FORMat? Query programmed math format
Return String:
NONE, MXB, or PERC
Description:
This command is used to specify the format for the CALC1 math calculations. With NONE
selected, no CALC1 calculation is performed. With MXB or PERCent selected and enabled, the
result of the calculation is displayed. The calculated reading is refreshed every time the instrument
takes a reading.
:KMATh Commands
:MMFactor <NRf>
Command syntax:
:CALCulate[1]:KMATh:MMFactor <NRf>
Command Parameter:
<NRf> = -100e6 to 100e6 Specify m factor.
Query:
:MMFactor? Query “m” factor.
Return String:
The value of m in <NRf> format
Description:
This command is used to define the “m” factor for the mx+b calculation.
:MBFactor <NRf>
Command syntax:
:CALCulate[1]:KMATh:MBFactor <NRf>
Command Parameter:
<NRf> = -100e6 to 100e6 Specify “b” factor.